Why Emitted Dose and Fine Particle Fraction Can Diverge

Illustration of a printed chromatogram beside a sample vial and pen

Written by

in

Article title illustration

Two inhalation powders can emit similar amounts of drug while producing different particle-size distributions. Tai and colleagues compared prototype CBD powders containing either human serum albumin or methyl-beta-cyclodextrin. They evaluated both emission and the fraction of aerosol in particles smaller than five micrometers.

Similar emission, different dispersibility

Both powders emitted about 90% of their CBD in the test. The fine particle fraction was approximately 30% for the albumin powder and 54% for the methyl-beta-cyclodextrin powder. The emitted fraction describes how much drug left the device. The fine particle fraction describes an aspect of the size distribution and provides additional information about aerosol behavior.

These are bench measurements. Actual deposition would depend on the device, the person’s breathing and airway conditions. Establishing how much CBD reaches the bloodstream would require a separate pharmacokinetic measurement.